|
 CDi Main | CDiComm | Controller | CDiNet | CDiBrowser | Outputs | Temp
| pH | DO | Wiring | Pricing
| |
| Keyboard Calibration |
Note: This section is provided for convenience only. CDi users are expected to download the latest help file
from the info section of this site.
The calibration process involves setting two points, minimum and maximum representing the input range of the variable.
The internal A/D (Analog to Digital) converter in the unit converts an analog voltage to counts, either in the
negative or positive direction. Note: You can switch the CDi to display either counts or translated values
by pressing the "V-Version" key. Press "V-Version"
again to toggle this function. The input range of the unit is from -1Volt to +1Volt, being internally represented
from -30000 to +30000 counts respectively. Most inputs only generate numbers in the positive part of the range.

Required tools
For temperature calibration you should have a thermometer.
For pH, you will need two different buffer solutions, preferably pH 4 and pH 10. If pH 10 buffer is not available,
use pH 7. For DO you should have 0.0ppm Oxygen solution. If not, you can "short" pins 2 and 3 on the
CPC connector to simulate zero. Do this only for DO channels.
Important!
Always wait for the reading to stabilize before before you accept the value indicated. The easiest
way to do this is to switch the unit to "counts" display (press the "V" key) and wait until
the reading is stable.
Keyboard Calibration
In the graphic identify the following:
1. Gray colored row of keys - these add 1000,
100, 10, 1, 0.1, 0.01 to the displayed value (F=+1000, K=+0.01)
2. Green colored row of keys - these subtract
1000,100, 10, 1, 0.1, 0.01 to the displayed value (L=-1000, Q=-0.01)
3. Arrow keys
4. Cal Min and Cal Max
keys
5. ENTER key |
| pH Calibration |
|
Calibration of pH channels requires buffer of pH 4.00 and 10.00. If buffer pH 10.00 is not available, use buffer
pH 7.00. For best temperature response both buffers should be at a temperature close to that of the water. Future
CDi software will auto-compensate for temperature.
To calibrate the minimum for a pH channel proceed as follows:
- Assuming you intend to calibrate the minimum for pH channel 2, expose the pH probe from channel 2 to pH 4.00
buffer.
Wait
for the readings to stabilize. You may wish to switch the unit to "counts" display by pressing the V-Version key. Watch the reading (2nd line on the left) change.
Do not proceed to the next step until the reading has stabilized.
- Press the Cal Min key, then left or right arrow keys to navigate to the channel
you want to calibrate (channel 2 in this example). If the indicated value is not 4.00, use the keys in the gray
row to add, or the keys in the green row to subtract.
- Press the ENTER key to accept or ESCAPE to
exit.
To calibrate the maximum for a pH channel proceed as follows:
- Assuming you intend to calibrate the maximum for pH channel 2, expose the pH probe from channel 2 to pH 10.00
buffer.
Wait
for the readings to stabilize. You may wish to switch the unit to "counts" display by pressing the V-Version
key. Watch the reading (2nd line on the left) change. Do not proceed to the next step until the reading has stabilized.
- Press the Cal Max key, then left or right arrow keys to navigate to the channel
you want to calibrate (channel 2 in this example). If the indicated value is not 10.00, use the keys in the gray
row to add, or the keys in the green row to subtract.
- Press the ENTER key to accept or ESCAPE to
exit.
|
| DO Calibration |
|
Calibration of DO channels requires familiarity with DO probe air calibration procedures. Fresh water saturation
data is given in this table. Look up the ambient temperature and match with
the pressure column. At sea level and 21 degrees C, the DO value is 8.9 mg/L. In this example we calibrate DO channel
4, assuming a DO saturation value of 8.9 mg/L.
To calibrate the minimum (zero) for a DO channel proceed as follows:
- Short the input (pins 2&3) on CPC connector for channel 4 with a shorting block.
Press
the Cal Min key, then left or right arrow keys to navigate to the channel you
want to calibrate (channel 4 in this example). If the indicated value is not 0.00, use the keys in the gray row
to add, or the keys in the green row to subtract.
- Press the ENTER key to accept or ESCAPE to
exit.
- Remove the shorting block and replace the original CPC connector.
To calibrate the maximum for a DO channel proceed as follows:
- Assuming you intend to calibrate the maximum for DO channel 4, fix the probe from channel 4 about 1cm above
the water surface and allow about an hour for temperature equalization.
Wait
for the readings to stabilize. You may wish to switch the unit to "counts" display by pressing the V-Version key. Watch the reading (4th line on the left) change. Do not proceed to the next
step until the reading has stabilized.
- Press the Cal Max key, then left or right arrow keys to navigate to the channel
you want to calibrate (channel 4 in this example). If the indicated value is not 8.90, use the keys in the gray
row to add, or the keys in the green row to subtract.
- Press the ENTER key to accept or ESCAPE to
exit.
|
| Temperature Calibration |
|
Calibration of temperature channels requires the use of a thermometer to verify actual temperature.
To calibrate the minimum for a temperature channel proceed as follows:
- Assuming you intend to calibrate channel 1, expose the temperature probe from channel 1 to a low temperature
(some ice in a bucket of water will do).
Wait
for the readings to stabilize. You may wish to switch the unit to "counts" display by pressing the V-Version key. Watch the reading (1st line on the left) change. Do not proceed to the next
step until the reading has stabilized. Monitor the temperature with the thermometer.
- Press the Cal Min key, then left or right arrow keys to navigate to the channel
you want to calibrate (channel 1 in this example). If the indicated value is not the same as the temperature indicated
on the thermometer, use the keys in the gray row to add, or the keys in the green row to subtract.
- Press the ENTER key to accept or ESCAPE to
exit.
To calibrate the maximum for a temperature channel proceed as follows:
- Assuming you intend to calibrate channel 1, expose the temperature probe from channel 1 to a high temperature
(place probe in usual position in tank. ~30 degrees C).
Wait
for the readings to stabilize. You may wish to switch the unit to "counts" display by pressing the V-Version key. Watch the reading (1st line on the left) change. Do not proceed to the next
step until the reading has stabilized. Monitor the temperature with the thermometer.
- Press the Cal Max key, then left or right arrow keys to navigate to the channel
you want to calibrate (channel 1 in this example). If the indicated value is not the same as the temperature indicated
on the thermometer, use the keys in the gray row to add, or the keys in the green row to subtract.
- Press the ENTER key to accept or ESCAPE to
exit.
|